How to Adjust the Pressure Settings on a Sullivan V Elite CPAP Machine
Things You'll Need
- Manometer (pressure measuring meter with a range of 0-22
- cm H2O in intervals of 0.1 cm)
- 6-foot length of standard oxygen tubing
- Male Luer taper fitting
- Blind bubble cushion (recommended)
Instructions
-
-
1
Plug your Sullivan V Elite CPAP machine into a regular wall outlet.
-
2
Set up the Sullivan V Elite with all of the accessories that you or the patient for whom you care uses; make sure to use the same mask that's used for actual treatment.
-
3
Flip the main power switch in the rear of the unit on and allow the machine to warm up for 15 minutes; press the stop button if the motor begins to run.
-
4
Insert the Luer taper into one end of the tubing and attach the other end to the manometer.
-
5
Remove the plug from one of the ports on the frame of your mask and insert the Luer taper end of the tubing into the port.
-
6
Turn the CPAP machine off.
-
7
Turn the main power switch on again while holding down the "Start," "5 minute" and "10 minute" buttons on the front console until all three buttons light up.
-
8
Block the flow of air to the mask, either with your hand or by attaching a blind bubble. Note that a blind bubble will not work with a "Mirage" type mask.
-
9
Press the "Stop" and "20 minutes" buttons together and then release them to start the motor and build up pressure.
-
10
Decrease the pressure as necessary by using the "5 minute" button or increase it with the "10 minute" button until the manometer reads "16 +/- 0.2 cm H20" for normal use or any other reading prescribed by your doctor or respiratory therapist.
-
11
Turn your CPAP machine off with the main power switch or press the "Stop" button to put it in regular use mode.
-
1