How to Set PS on BiPAP Auto
Things You'll Need
- Respironics BiPAP Auto M Series CPAP device
- Power cord
Instructions
-
-
1
Plug the small round plug on one end of the power cord into the power jack on the CPAP machine.
-
2
Press and hold the right and left arrow keys on the control pad while plugging the other end of the power cord into a wall outlet. Then press the "+" button on the control pad. This accesses the clinician's setup menu.
-
3
Press the right arrow key until you see the IPAP screen. This shows the positive airway pressure provided during inhalation, measured in centimeters of water or cmH20. Press the "+" or "-" button to set this to the prescription pressure.
-
4
Press the right arrow key until you see the EPAP screen. EPAP is the amount of air pressure provided during the exhalation phase. Use the "+" or "-" buttons to adjust this setting.
-
5
Press the right arrow key until you see the MAX IPAP screen. This is the maximum air pressure the machine will deliver if the BiPAP feature is enabled. Use the "+" or "-" buttons to adjust this setting.
-
6
Press the power button to exit the clinician mode and return to user mode.
-