Medtronic Troubleshooting
Instructions
-
-
1
Inspect the electrodes if electrode impedance appears too high. Re-insert the dislodged electrode into the patient, and tape it in place if it was partially dislodged. If it was completely dislodged, replace the electrode. Replace a high-resistance electrode with a new electrode. Check the connection at the patient interface box to verify that the electrode pins are firmly joined.
-
2
Remove and replace an electrode that is dirty or has a broken cable. A dirty electrode or broken cable can cause the electrode to read "OFF." Damaged cables can have a similar effect.
-
3
Change the muting probe connector to a lower setting until it stops muting. If it still mutes on the lowest setting, disconnect the muting detector entirely. The muting detector must not interfere with necessary systems operation.
-
1